Basic Ephys Data Processing in Bonsai

This tutorial shows how to use ONIX hardware and the OpenEphys.Onix1 Bonsai package to perform basic online signal processing on electrophysiology data in Bonsai such as channel selection and reordering, frequency filtering and event detection (in this example, spike detection using a fixed threshold crossing).

This type of processing is helpful for visualizing data during acquisition and can be a starting point for more advanced workflows such as closed-loop experiments. For specialized data visualizations from very dense arrays like Neuropixels probes, for example, we recommend piping that data to the Open Ephys GUI.

This tutorial guides you through building the following workflow:

/workflows/tutorials/basic-ephys-processing/spikes.bonsai workflow

Note

Although this tutorial uses headstage64 as an example, the process is similar for other ephys headstages. This tutorial assumes you are familiar with the hardware guide of the ONIX headstage you intend to use. Use this reference for which ephys Data I/O Operators and scaling you need to use for each headstage, and links to relevant documentation.

Set up and get started in Bonsai

Follow the Getting Started guide to set up and get familiarized with Bonsai. In particular:

Configure the hardware

/workflows/tutorials/basic-ephys-processing/configuration.bonsai workflow

Construct a top-level hardware configuration chain:

  1. Place the configuration operators that correspond to the hardware you intend to use between CreateContext and StartAcquisition. In this example, these are ConfigureHeadstage64 and ConfigureBreakoutBoard.
  2. Confirm that the device that streams electrophysiology data is enabled. The Rhd2164 device (an Intan amplifier) on the headstage64 is the only device used in this tutorial, so you could disable other devices on the headstage and on the breakout board to improve performance if you wanted to.

Stream ephys data into Bonsai

/workflows/tutorials/basic-ephys-processing/ephys-data.bonsai workflow

Place the relevant operators to stream electrophysiology data from your headstage:

  1. Because the device on headstage64 that streams electrophysiology data is the Rhd2164 Intan amplifier, we placed the Rhd2164Data node onto the workflow. Use this reference to find the ephys data operator that corresponds to each device.
  2. Select the relevant members from the data frames that the data operator produces. In this example, the relevant members are "AmplifierData" and "Clock". To select those members, right-click the Rhd2164 node, hover over the output option in the context menu, and select it from the list.
  3. Visualize the raw data to confirm that the ephys data operator is streaming data.

Select and reorder channels

/workflows/tutorials/basic-ephys-processing/select-convert-ephys-data.bonsai workflow

Connect a SelectChannels operator to the electrophysiology data stream and edit its "Channels" property.

  • Remember indexing in Bonsai starts at 0.
  • Use commas to list multiple channels and brackets for ranges.
  • Reorder channels by listing the channel numbers in the order in which you want to visualize the channels.

Convert ephys data to microvolts

/workflows/tutorials/basic-ephys-processing/select-convert-ephys-data.bonsai workflow

Center the signal around zero

Connect a ConvertScale operator to the SelectChannels operator and set its properties:

  • Edit its "Shift" property to subtract 2bit depth - 1 from the signal. Use this reference to find the Shift necessary for each device. In this example, we "Shift" -32768 because the Rhd2164 device outputs unsigned 16-bit data.
  • Set the "Depth" property to F32 because this bit depth is required to correctly represent scaled data from all devices.

Scale the signal to microvolts

Connect a second ConvertScale operator to the first ConvertScale operator and set its properties:

  • Edit its "Scale" property to multiply the signal by a scalar in order to get microvolt values. This scalar is determined by the gain of the amplifier and resolution the ADC contained in the amplifier device. Use this reference to find the "Scale" necessary for each device. In this example, we "Scale" by 0.195 because the Rhd2164 device on headstage64 has a step size of 0.195 μV/bit
  • Keep the "Depth" property at F32.

Visualize the transformed data to confirm the output of the shifting and scaling operations worked as expected, i.e. that the signal is centered around zero and that the values make sense in microvolts.

Note

Although both the Shift and Scale calculation can be done in one ConvertScale operator, the calculations are more straightforward using two operators connected in series because the ConvertScale operator applies the "Shift" offset after applying the "Scale" scalar so if we used a single operator, we would have to scale the Shift parameter.

Apply a filter

/workflows/tutorials/basic-ephys-processing/filter-ephys-data.bonsai workflow

Connect a FrequencyFilter operator to the second ConvertScale operator and set its properties.

  • Set its "SampleRate" property to 30000. Ephys data in all devices is 30 kHz.
  • Set the "FilterType" property to an adequate type. In this example, we use a high pass filter to look at spikes.
  • Set the "Cutoff1" and "Cutoff2" properties to an adequate value. In this example, we use 300 Hz as the lower cutoff frequency.

Visualize the filtered data.

Tip

If you choose to save data, we recommend you place the MatrixWriter operator before filtering and scaling to save raw data instead of scaled or filtered data. Filtering with the FrequencyFilter operator before recording could remove signals from a bandwidth of interest and converting to microvolts with the second ConvertScale operator could increase the size of your data without increasing meaningful information.

Detect events

/workflows/tutorials/basic-ephys-processing/spike-detection.bonsai workflow

Based on the amplitude of the signal on the selected channel, set a fixed threshold for detecting spikes.

Visualize the spike data.

Tip

You can test the spike detection using a pre-recorded data known to have spikes: recreate the workflow from this example without the hardware configuration chain in a new workflow and replace the ephys data node (in the case of the headstage64, replace the Rhd2164 node) with a MatrixReader that reads from the file containing spiking ephys data in unsigned 16-bit format.