Process and Listen to Ephys Data

This tutorial shows you how to perform basic online signal processing of electrophysiology data in Bonsai such as channel selection/reordering, frequency filtering, and fixed-threshold spike detection as well as how to listen to ephys data using ONIX hardware and the OpenEphys.Onix1 Bonsai package.

Note

This tutorial serves primarily as an introduction to basic operations for processing electrophysiology data in Bonsai. These operations can also be performed in the Open Ephys GUI which provides advanced visualizations and built-in processing modules.

/workflows/tutorials/ephys-process-listen/ephys-process-listen.bonsai workflow

Tip

This tutorial uses the Headstage 64. To find pertinent information for processing data produced by the Intan Rhd2164 ephys device on the Headstage 64 (such as the numbers used to convert DAC data to units of voltage), navigate to Rhd2164DataFrame which is the page for the data element produced by the Rhd2164Data operator. The process of navigating the reference to find pertinent information is the same for other ephys headstages that you might want to use to follow along in this tutorial. It also helps to be familiar with the example workflow for your particular hardware available in the Data Acquisition Quick Start.

Get Started in Bonsai

Follow the Getting Started guide to set up and familiarize yourself with Bonsai. In particular:

Configure the Hardware

Construct a top-level hardware configuration chain:

/workflows/tutorials/ephys-process-listen/configure.bonsai workflow

  1. Place the configuration operators that corresponds to the hardware you intend to use between CreateContext and StartAcquisition. In this example, these are ConfigureBreakoutBoard and ConfigureHeadstage64.
  2. Confirm that the device that streams electrophysiology data is enabled. The Rhd2164 device (an Intan ephys acquisition acquisition chip) on the headstage 64 is the only device used in this tutorial, so you can disable other devices on the headstage and on the breakout board.

Stream Ephys Data into Bonsai

Place the relevant operators to stream electrophysiology data from your headstage:

/workflows/tutorials/ephys-process-listen/ephys-data.bonsai workflow

  1. We placed the Rhd2164Data place into the workflow because the device on headstage 64 that streams electrophysiology data is the Rhd2164 Intan amplifier.
  2. Select the relevant member from the data frames that Rhd2164Data produces. In this example, the relevant member is AmplifierData. To do this, right-click Rhd2164Data, hover over the output option in the context menu, and select "AmplifierData" from the list.

Visualize the raw data to confirm that the ephys data operator is streaming data.

Detect Spikes

Select and reorder channels

Connect a SelectChannels operator to the electrophysiology data stream and edit its "Channels" property:

/workflows/tutorials/ephys-process-listen/spike_select-channels.bonsai workflow

  • Remember indexing starts at 0.
  • Reorder channels by listing the channel numbers in the order in which you want to visualize the channels.

Center the signal around zero

Connect a ConvertScale operator to the SelectChannels operator and set its properties:

/workflows/tutorials/ephys-process-listen/spike_center-data.bonsai workflow

  • Edit its Shift property to subtract 2bit depth - 1 from the signal. In this example, we shift -32768 because the Rhd2164 device outputs unsigned 16-bit data.
  • Set the Depth property to S16 or F32. A sufficiently large data type is required to represent ephys data without overflow.

Scale the signal to microvolts

Connect a second ConvertScale to the first ConvertScale and set its properties:

/workflows/tutorials/ephys-process-listen/spike_scale-data.bonsai workflow

  • Edit its Scale property to multiply the signal by a scalar in order to get microvolt values. This scalar is determined by the gain of the amplifier and resolution the ADC contained in the bioacquisition device. In this example, we scale by 0.195 because the Rhd2164 device on headstage64 has a step size of 0.195 μV/bit.
  • Set the Depth property at F32 which is required to represent decimal values.

Visualize the transformed data to confirm the output of the shifting and scaling operations are performed as expected, i.e. that the signal is centered around zero and that the values make sense in microvolts.

Note

Although both the shift and scale computations can be done with a single ConvertScale, the calculation is more straightforward using two operators connected in series because ConvertScale applies the "shift" transformation after applying the "scale" transformation. If we used a single operator, we would have to pre-scale the Shift property.

Apply a filter

Connect a FrequencyFilter operator to the second ConvertScale operator and set its properties:

/workflows/tutorials/ephys-process-listen/spike_filter-data.bonsai workflow

  • Set its SampleRate property to 30000. Ephys data in all devices is 30 kHz.
  • Set the FilterType property to an adequate type. In this example, we use a band pass filter to look at spikes and remove slowly changing signals.
  • Set the Cutoff1 and Cutoff2 properties to adequate values.

Visualize the filtered data to confirm that it matches your expectations.

Tip

If you choose to save data, we recommend you place the MatrixWriter operator before filtering and scaling to save raw data instead of scaled or filtered data. Converting to microvolts with the second ConvertScale operator increases the size of your data (because it's converted to F32 from S16) without increasing meaningful information. Filtering with the FrequencyFilter operator before recording removes signal in irrecoverable ways. It's preferable to either save both or apply another filter when loading and processing the data.

Detect spikes with fixed threshold

/workflows/tutorials/ephys-process-listen/spike_detect.bonsai workflow

Based on the amplitude of the signal on the selected channel, set a fixed threshold for detecting spikes.

Visualize the spike data.

Listen to Ephys

The output of AmplifierData can be directed into two separate signal processing streams. In other words, it is possible for two downstream operators to receive the same sequence of AmplifierDataFrames. This is helpful for creating two distinct disparate processes for the same data stream, one for visualizing spikes in several channels as we did above and one for listening to a single channel as follows.

Select a channel and process the signal for audio

/workflows/tutorials/ephys-process-listen/audio_process.bonsai workflow

The same basic steps as the Spike Detection section are performed. However, the property settings are critically different.

  • Select a single channel for listening instead of multiple channels for detecting spikes.
  • Use the second ConvertScale to scale the signal by a value between 0 and 1 to control the volume of the audio signal instead of scaling by 0.195 to convert to microvolts. This ConvertScale effectively serves as a volume knob.

Visualize the data and compare it at various points in the processing pipeline to confirm it matches your expectations.

Play ephys data as audio

/workflows/tutorials/ephys-process-listen/ephys-process-listen.bonsai workflow

Connect an AudioPlayback operator to FrequencyFilter and set its SampleRate property to 30000.

Providing data to AudioPlayback that is outside of the bounds of a signed 16 bit integer (-32,768 to +32,767) introduces clipping distortion into the audio signal, so it is recommended to maintain a Scale property value of less than 1 for the volume knob ConvertScale. Maximize other volume settings on your PC before exceeding 1.