Headstage Rhs2116
These are the devices available on the Headstage Rhs2116:
- Rhs2116 Pair:
- 32 electrophysiology channels that are independently configurable to record electrophysiological signals from or
deliver electrical stimuli to passive probes.
- 30193.2367 Hz sample rate with 16 bit depth
- Adjustable analog band-pass filter:
- Lower cutoff configurable from 0.1 Hz to 1 kHz
- Upper cutoff configurable from 100 Hz to 20 kHz
- Lower cutoff during recovery immediately after stimulus configurable from 0.1 Hz to 1 kHz
- Optional adjustable digital high-pass filter with cutoff configurable from 0.146 Hz to 3.309 kHz
- 32 electrophysiology channels that are independently configurable to record electrophysiological signals from or
deliver electrical stimuli to passive probes.
- Electrical Stimulation: Single current source with ±15V compliance voltage and automatic electrode discharge
- The stimulation waveform is highly configurable via the Rhs2116StimulusTrigger's properties.
Tip
Visit the Headstage Rhs2116 Hardware Guide to learn more about the hardware such as weight, dimensions, and proper power voltages.
The following pages in the Headstage Rhs2116 Guide provide a breakdown of the above example workflow.