Deprecated since version 1.2.13: Bonsai.ONIX is deprecated.

To use ONIX with Bonsai, refer to the documentation for OpenEphys.Onix1.

MiniscopeV4Device#

Important

To use Miniscopes with ONIX hardware, you must first use the ONIContext configuration GUI to set the headstage port to “Passthrough Mode”

A Bonsai source that wraps a DS90UB9X Raw Device device and configures it to allow control over UCLA Miniscope V4.

Inputs:

None

Outputs:

A single MiniscopeV4DataFrame that is produced periodically by the scope and contains an image and corresponding sample and frame counters.

UCLAMiniscopeV4.bonsai
UCLAMiniscopeV4

Configuration#

Configuration is performed using the property pane which contains the following options.

Name

Type

Description

EnableStream

boolean

Enable the device data stream

LEDBrightness

double

Excitation LED brightness (0-100%).

InterleaveLED

bool

If true, the excitation LED will only turn on during camera exposures. This can reduce photobleaching and tissue heating.

Gain

SensorGain (enum)

Camera sensor analog gain

FrameRate

FPS (enum)

Camera images acquired per second.

LiquidLensVoltage

double

The RMS voltage output by the electrowetting lens driver which determines the lens focal length